The industry standard Hach CL17 Chlorine Analyzer uses colorimetric DPD chemistry to continuously monitor water for free or total chlorine. For typical use, the CL17 will operate unattended for 30 days and is compliant with US EPA regulation 40 CFR 140.74. Monthly routine maintenance for the CL17 can be performed in 15 minutes, including changing reagents and cleaning the colorimetric cell. No special tools are required.
The CL17 analysis method is not affected by changes in chlorine concentration, sample pH, temperature, flow, or pressure, thus offering more reliable and accurate measurement with less operator interventions than other methods in today's market.
The CL17 is also suitable for permanganate residual measurements. Laboratory and field studies demonstrate the analytical capabilities of the analyzer to determine permanganate concentration in water and confirm that it is a robust solution and a good tool for the control of challenging water treatment processes (variable water quality and demand).
Accessories: Installation Kit Maintenance Kit
Accuracy: ± 5 % or ±0.03 mg/L as CL₂, whichever is greater
Air Purge: 0.1 cfm (0.17 m³/h) instrument quality air at max. 20 psig (ca. 1.4 bar) with 1/4-inch OD tube, quick disconnect fitting
Alarm: Two alarms selectable for sample concentration alarm, analyzer system warning, or analyzer system shut-down alarm. Each is equipped with an SPDT relay with contacts rated for 5A resistive load at 230 V AC.
Certifications:
Europe, CE Approved with:
EN 61326-1
CISPR 11
EN 61010-1
IEC 60529
North America: UL 61010A-1
CAN/CSA C22.2 No. 1010.1-92
Communication Capabilities: 4-20mA Output
Cycle Time: 2.5 min
Description : CL17 Chlorine Analyzer with Reagents for Free Chlorine
Dimensions (H x W x D): 454 mm x 314 mm x 179 mm
Display: LCD, 3-1/2 inch digit measurement readout and six-character alphanumeric scrolling text line
Drain: 1/2-inch ID flexible hose, hose barb
Enclosure Construction: ABS plastic, two clear polycarbonate windows, IP62-rated with the gasketed door latched.
Enclosure Rating: IP62
Includes: CL17 Chlorine Analyser with 1 Month Supply of Free Chlorine Reagents, Maintenance kit, Installation Kit, and Manual. (The power cord is ordered separately.)
Inlet: 1/4-inch OD polyethylene tube, quick-disconnect fitting
Inlet Pressure: 1 to 5 psi (0.07 to 0.34 bar), 1.5 psi (1.03 bar) is optimum
Interferences: Other oxidizing agents such as bromide, chlorine dioxide, permanganate and ozone will cause a positive interference. Hexavalent chromium will cause a positive interference: 1 mg/L Cr6+ = approximately 0.02 mg/L as Cl₂. Hardness must not exceed 1,000 mg/L as CaCO₃.
Light Source: Class 1 LED (light emitting diode) with a peak wavelength of 520 nm; 50,000 hours estimated minimum life
Lower Limit of Detection (LOD): 0.03 mg/L (ppm)
Mounting: Wall mount
Operating Humidity: Up to 90 % at 40 °C (104 °F) maximum
Operating Temperature Range: 5 - 40 °C max.
Parameter: Chlorine, Free & Total
Power Requirements (Amps): 2.5 A
Power Requirements (Hz): 50/60 Hz
Power Requirements (Voltage): 100 - 115/230 V AC
Pressure Range: Inlet Pressure to Sample Conditioning: 1.5 - 75 psi (0.1 - 5.2 bar)
Range: 0 - 5 mg/L Cl₂
Recorder Outputs: One 0/4-20 mA with an output span programmable over any portion of the 0 to 5 mg/L range. Recommended load impedance 3.6 to 500 ohms, 130 V isolation from earth ground.
Repeatability: 0.01 mg/L or 3.0 % , whichever is greater
Sample Flow Rate: 200 - 500 mL per minute minimum
Sample Temperature: 5 - 40 °C (41 - 104 °F)